0 etco2 monitor test (for etco2 option), Etco2 monitor test (for etco2 option) -33, 0 etco – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 43: Monitor test (for etco, Option)

Advertising
background image

R Series Service Manual

9650-0903-01 Rev. J

1-

33

19.0 EtCO

2

Monitor Test (for EtCO

2

Option)

Tools Needed: CAPNOSTAT 5 Mainstream cable with airway adapter.

Test Setup:

1. Install the battery.

Do this...

Observe this...

Pass/Fail

19.1

Connect the CAPNOSTAT 5 CO2 Mainstream
cable with airway adapter attached to the yellow
connector at the back of the R Series.

NOTE: Make sure the airway adapter is installed in the CO

2

cable.

19.2

Set the front panel switch to MONITOR or ON.
For AED units, enter Manual Mode.

WARM UP message appears on the display.

NOTE: Warming up may take about 1 minute.

o

o

19.3

When the WARM UP message disappears,
press the Param softkey, then select EtCO2
and press Enter.

19.4

Press the ZERO softkey, then wait for the
ZERO DONE message.

The ZERO DONE message appears.

o

o

19.5

Press the Return softkey.

19.6

Press the Wave 2 softkey until the CO

2

waveform appears.

A flat baseline CO

2

waveform appears.

o

o

19.7

Breath normally into the airway adapter.

A capnogram waveform appears.

o

o

Record your results on the Maintenance Test Checklist.