Code readiness log – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 313

Manual Defibrillator Testing
9650-0904-01 Rev. K
ZOLL R Series Operator’s Guide
13–7
•
10 mm ±1.0 mm high
Code Readiness Log
Each automatic and manual defibrillation test result is stored in internal non-volatile memory,
the Code Log. A total of 1000 Code Readiness test records can be stored in internal memory.
When the Code Readiness Log is full, the oldest records are erased on a first-in-first-out (FIFO)
basis.
The Code Readiness Log can be transferred to an external computing device (see “Transferring
Files to an External Device” on page 12-1).
If configured, the R Series prints a Code Readiness Test Report following the completion of
each automatic defibrillator test.
To print the Code Readiness Log:
1. Turn the Mode Selector to ON.
2.
Press the Manual Mode softkey.
Note:
If the CONFIRM MANUAL MODE message displays, press the Confirm softkey to
enter the Manual mode of operation.
3.
Press the Report Data softkey.
4.
Press the Test Log softkey.
A menu appears with the print options.