Transferring files through the usb port (optional) – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 476

Advertising
background image

C

HAPTER

11

F

ILE

T

RANSFER

11–4

www.zoll.com

9650-0912-01 Rev. K

2.

Press the Report Data softkey.

3.

Press the Transfer Mode softkey.

4.

Press the More softkey.

5.

Press the Defib History to Card softkey.

The message TRANSFERRING DATA is displayed. All data is transferred to the installed CF
data card.

Note:

Do not remove the CF card while files are transferring. Corruption may result on the
data card.

When all files are transferred, the message DATA TRANSFERRED is displayed. You can now
remove the CF card from the R Series unit.

To exit the Data Transfer mode, press the Return softkey and then the Exit Transfer softkey.

Transferring Files Through the USB Port (Optional)

Before you begin, connect a USB cable from the Windows external device with USB host
capability (for example, a Window PC), to the R Series defibrillator USB device port.

Figure 11-3.

USB Ports

To transfer data through the USB port:

1. Turn the Mode Selector to MONITOR.

2.

Press the Report Data softkey.

3.

Press the Transfer Mode softkey.

4.

Press the More softkey.

5.

Press the Enable USB softkey.

The R Series is now in USB Transfer Mode.

6.

Initiate data transfer using ZOLL CodeNet Central.

To exit USB Transfer mode, press the Disable USB softkey or switch the Mode Selector to
DEFIB, PACER or OFF.

USB Device Port

USB Host Port
(Future Use)