Removing the system brick assembly, Removing the system brick assembly -23 – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 89
Advertising
R Series Service Manual
9650-0903-01 Rev. J
3-23
Removing the System Brick Assembly
Note:
See the safety precautions on 3-2 before removing the system brick assembly.
The system brick assembly consists of three boards attached together.
Top: Digital system board
Middle: Analog system board
Bottom: Pace defibrillator engine
Caution
Use caution when pulling apart the three boards to prevent the EMI suppression plates from scattering.
1. Remove the printer housing barrier from the chassis.
2. Remove the ECG retainer pad.
Note:
During reassembly, the barrier may be reused if the adhesive
side is not contaminated.
Note:
When reinserting the pad during reassembly, push it down until
it touches the ECG cable connector.