Full disclosure recording, Incident logs, Printing a partial summary report – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 470: Printing an incident log

Advertising
background image

C

HAPTER

10

E

VENT

R

ECORDS

AND

R

EPORTS

10–8

www.zoll.com

9650-0912-01 Rev. K

Printing a Partial Summary Report

To print a partial summary report:

1. Press the Report Data softkey, then press Print Chart.

2.

Press the Print Range softkey.

3.

Use the First Event, Prev. Event, Next Event and Last Event softkeys to locate the event
from which printing will start (all subsequent events are also printed).

4.

Press the Print softkey.

The unit prints all records from the selected item to the most recent.

Note:

The overview information (such as the number of shocks delivered) covers all stored
summary data, not just the selected range.

Full Disclosure Recording

Along with event information captured in Summary Report, R Series also records the full
disclosure CPR sensor and parameter waveforms. Full disclosure recordings are erased at the
same times as Summary Reports.

Incident Logs

An incident log is an abbreviated list of all major events recorded in summary report. You can
print an incident log that includes the following events and their time of occurrence.

Unit powered on.

Defibrillation advisory message issued (for example, CHECK PATIENT or SHOCK
ADVISED
)

Shock delivered (and energy level)

PACER mode selected

Alarm triggered

Stripchart printing started

Code marker entered

In addition, the incident log lists the following:

System serial number

Device identification number

Report start time (when the summary data was last erased)

Time of the last recorded event

Total number of shocks delivered

Total cumulative pacing time

Printing an Incident Log

To print an incident log, press the Report Data softkey, then press the Print Log softkey.

The log is printed on the stripchart, starting with the oldest entry.