R Series Service Manual
9650-0903-01 Rev. J
1-
17
8.0 Calibrating Pulses on Strip Chart Test
Tools Needed: None.
Test Setup: None
.
Do this...
Observe this...
Pass/Fail/NA
8.1
Press the RECORDER button.
8.2
Press and hold SIZE button to activate the
calibration signal.
The strip chart displays a signal of 300 ppm with an amplitude of 10 mm
+/- 1 mm. The signal also appears on the video display.
o
o
o
Record your results on the Maintenance Test Checklist.