R Series Service Manual
1-14
9650-0903-01 Rev. J
5.0 Leakage Current Test
Tools Needed: See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Test Setup: See the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with applicable accessories:
OneStep Cable, external paddles, internal paddles, and anterior/posterior paddles.
Maximum Leakage Acceptance Limits
Normal Condition
Single Fault Condition*
Lead ECG
10
50
Hands Free Electrodes and Paddles (MFE)
10
100
Earth
500
1000
*Single fault considered AC mains on applied part.