R Series Service Manual

1-22

9650-0903-01 Rev. J

11.3

Select energy level of 100J and press the
CHARGE button.

The charge time is >2 second and <10 seconds and SELECT 30J FOR TEST
is displayed.

o

o

11.4

Press the SHOCK button.

Unit does not discharge. DEFIB PAD SHORT message displays.

o

o

11.5

Set energy level to 30J.

Unit internally discharges.

o

o

11.6

Press the CHARGE button.

Unit charges to 30J and displays DEFIB 30J READY. The charge ready tone
sounds.

o

o

11.7

Press and hold SHOCK button.

Unit discharges. 30J TEST OK message displays. The message at the top of
the printed strip chart reads as follows:

30 JOULES TEST OK. TEST_CUR=10-14A DEFIB_IMPED=0.

The impedance value may range from 0 to 5

.

o

o

Record your results on the Maintenance Test Checklist.

Do this...

Observe this...

Pass/Fail