ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 32

R Series Service Manual
1-22
9650-0903-01 Rev. J
11.3
Select energy level of 100J and press the
CHARGE button.
The charge time is >2 second and <10 seconds and SELECT 30J FOR TEST
is displayed.
o
o
11.4
Press the SHOCK button.
Unit does not discharge. DEFIB PAD SHORT message displays.
o
o
11.5
Set energy level to 30J.
Unit internally discharges.
o
o
11.6
Press the CHARGE button.
Unit charges to 30J and displays DEFIB 30J READY. The charge ready tone
sounds.
o
o
11.7
Press and hold SHOCK button.
Unit discharges. 30J TEST OK message displays. The message at the top of
the printed strip chart reads as follows:
30 JOULES TEST OK. TEST_CUR=10-14A DEFIB_IMPED=0.
The impedance value may range from 0 to 5
.
o
o
Record your results on the Maintenance Test Checklist.
Do this...
Observe this...
Pass/Fail