Recorder

9650-0912-01 Rev. K

ZOLL R Series Operator’s Guide

13–3

Recorder

6. Noisy ECG, artifact, wandering

baseline

Consider 1 – 21Hz filter bandwidth (see R Series
Configuration Manual
).

Prepare the patient’s skin prior to electrode attachment.

Check for proper adhesion of electrodes to patient.

Reduce or eliminate ECG artifact due to electrode or
patient cable movement. Route cables so that they don’t
pull on electrodes or swing excessively.

Ensure patient is motionless.

Check for possible excessive radio frequency
interference.

7. Poor ECG signal amplitude,

calibration pulse normal

Select another lead.

Apply new electrodes using different placement.

8. Inconsistent QRS beep or heart

rate

Select another lead.

Alter ECG electrode placement.

9. Sync marker is absent or

inconsistent with QRS waveform
on display and recorder

Ensure device is in Sync mode.

Ensure that device in NOT in Remote Sync mode.

Change ECG lead selection.

Alter ECG electrode placement.

Paper too narrow. It should be 90 mm wide.

10. No ECG trace or dashed line on

display

Device is in Remote Sync mode. Press the
Sync On/Off softkey to exit Remote Sync mode.

Symptom

Recommended Action

1. CHECK RECORDER message

Recorder out of paper.

Remove paper, check paper type, check recorder for
paper jam, and reload paper.

Recorder door is open.

Paper is loaded upside down.

2. Recorder makes stuttering sound

when activated.

Check recorder for paper jam.

3. Light or poor quality printing

Ensure correct paper is in use.

Ensure paper is installed grid side against recorder print
head.

Recorder print head requires cleaning.

4. Summary Report will not print

when Report/Print Chart softkeys
are activated.

15 seconds have not elapsed since one of the events
that trigger Summary Report to record have occurred.
Wait 15 seconds, then try again.

Symptom

Recommended Action