R Series Service Manual
3-2
9650-0903-01 Rev. J
•
Needlenose pliers
•
Exacto knife
•
Compressed air
•
1/2” Nut Driver
•
3M Scotch-Weld Hot Melt Adhesive 3779 TC Amber
Safety Precautions
Warning!
SHOCK HAZARD!
Caution
TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY BEFORE YOU CONDUCT
DEFIBRILLATOR TESTS OR REPAIRS.
•
Only properly trained technicians should service the unit.
•
The unit can contain deadly voltages even if the unit is turned off.
•
Make sure to discharge the unit before working with it.
•
Make sure you take the necessary precautions when working with static sensitive units. For example, you must wear a conductive wrist strap (which
touches your skin) connected to a grounding mat and to the earth ground. You must remove the wrist strap when you discharge high voltage or when
you are working on energized equipment.
•
Wear gloves to prevent skin oils from affecting the equipment.