0 power supply test (optional), Power supply test (optional) -9 – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 19

Advertising
background image

R Series Service Manual

9650-0903-01 Rev. J

1-

9

4.0 Power Supply Test (Optional)

Note:

Tests in this section will produce battery errors due to the use of a power supply in place of a SurePower Battery.

Tools Needed:

2 red miniature alligator to miniature alligator leads.

2 black miniature alligator to miniature alligator test leads.

DC power supply (15 Amp minimum).

0.1



resistor (

¼

W or greater).

1000

1%

¼

W resistor.

Fluke 75 multimeter or equivalent.

Test Setup:

1. Disconnect the AC line cord from the unit.

2.

Make sure the unit and power supply are turned off.

3.

Connect one end of the black lead to the “-” terminal in the battery well.

4.

Connect the other end of the black lead to the “-” terminal of the power supply.

5.

Connect the red lead to “+” terminal socket of the battery well. Use the middle pin with the plastic guard around it. Connect the other end of the red lead
to the “+” terminal of the power supply.

6.

Set the power supply voltage to 7V.

Caution

Be sure to connect the power supply properly to the R Series battery well terminals or damage to the unit may result. Do NOT raise the
power supply voltage above 15V.