0 power supply test (optional), Power supply test (optional) -9 – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 19

R Series Service Manual
9650-0903-01 Rev. J
1-
9
4.0 Power Supply Test (Optional)
Note:
Tests in this section will produce battery errors due to the use of a power supply in place of a SurePower Battery.
Tools Needed:
•
2 red miniature alligator to miniature alligator leads.
•
2 black miniature alligator to miniature alligator test leads.
•
DC power supply (15 Amp minimum).
•
0.1
resistor (
¼
W or greater).
•
1000
1%
¼
W resistor.
•
Fluke 75 multimeter or equivalent.
Test Setup:
1. Disconnect the AC line cord from the unit.
2.
Make sure the unit and power supply are turned off.
3.
Connect one end of the black lead to the “-” terminal in the battery well.
4.
Connect the other end of the black lead to the “-” terminal of the power supply.
5.
Connect the red lead to “+” terminal socket of the battery well. Use the middle pin with the plastic guard around it. Connect the other end of the red lead
to the “+” terminal of the power supply.
6.
Set the power supply voltage to 7V.
Caution
Be sure to connect the power supply properly to the R Series battery well terminals or damage to the unit may result. Do NOT raise the
power supply voltage above 15V.