ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 55

Advertising
background image

R Series Service Manual

9650-0903-01 Rev. J

2-3

CF TRANSFER FAILED

Summary/DVCK/ALOG data file transfer error – either no
CF card or CF card transfer failed

Reseat CF Communication cable or replace
Communication Module.

CHANGE LEADS

Unit is in Defib Sync mode and heart rate is less than 20
BPM.

CHARGE FAILED

Unit failed to perform the requested charge.

Replace PD Engineer or Analog Board.

CHECK CO2 ADAPTER

Airway adapter is removed, occluded or adapter zeroing
needs to be performed or was performed incorrectly.

Replace/Clean airway adapter. Zeroing performed
automatically.

CHECK CO2 SENSOR

EtCO

2

Sensor is unplugged or defective.

Check that sensor cable is plugged in and seated
properly. Check that sensor is not exposed to
excessive heat. If problem persists, replace the
sensor.

CHECK CUFF/HOSE

Blood pressure cuff or hose is not installed correctly.

Cuff or hose is faulty.

Hose is kinked or disconnected.

Inflation rate too fast or too slow.

Verify the hose and cuff is properly connected
and not leaking.

Replace cuff and hose.

Replace NIBP module or parameter power
supply.

CHECK ECG CABLE

Invalid ECG cable Id is detected.

Verify that the ECG Cable is property
connected.

Replace ECG Cable.

Replace Analog Board.

CHECK ELECTRODE

Either Can read ID Chip or the checksum failed.

Replace Electrodes or replace Analog board.

CHECK PADS

Message displayed in conjunction with either POOR PAD
CONTACT or DEFIB PAD SHORT.

Ensure pads are coupled to patient. Check/
replace pads and universal cable. Replace system
board.

CHECK PATIENT

Background ECG analysis detects shockable rhythm.

CHECK PULSE

Alternate message for NO SHOCK ADV. message, or
displayed in addition to it according to the configuration
option selected. Also shown after delivering last shock
when Auto Analyze option is enabled.

Error Message

Explanation

User
Advisory

Technical Action