R Series Service Manual

2-10

9650-0903-01 Rev. J

POOR PAD CONTACT

Electrode impedance exceeds threshold.

Ensure pads are coupled to patient.

Check/replace pads or universal cable.

Check impedance circuit calibration.

Replace system board.

PRESS ANALYZE

Alternate message for Check Patient message if this
configuration option is selected.

PRESS CHARGE

Discharge button is pressed in Manual Defib mode but
the unit is not charged.

Advisory message following SHOCK ADVISED in
Manual/Advisory Defib mode with auto-charge
disabled.

PRESS SHOCK

Prompt issued in AED auto defib mode when defib is
charged (ready).

READING LOG DATA

Summary Report Log data printing message.

RECORDER FAULT 142

Strip chart system error.

Replace printer assembly, digital board, and/or
printer interconnect board.

RECORDER FAULT 143

Strip chart failed power-up echo test.

Replace printer assembly, digital board, and/or
printer interconnect board.

RECORDER FAULT 147

Strip chart printhead over safe operating temperature.

Replace printer assembly, digital board, and/or
printer interconnect board.

RELEASE BUTTONS

Simultaneous external paddle button presses detected
before unit reached full defib charge (ready state).

Release buttons.

RELEASE SHOCK

Discharge switch(es) closed when pressing charge
button.

Discharge button pressed before defib reached ready
state.

Release shock button.

Check paddles.

Replace controls board.

REMOVE SYNC

Analyze button pressed, or no heart rate detected while in
Defib Sync mode.

Error Message

Explanation

User
Advisory

Technical Action