9650-0912-01 Rev. K

ZOLL R Series Operator’s Guide

10–1

Chapter 10

Event Records and Reports

The R Series defibrillator records important event information during operation. You can
retrieve this information in various forms:

Summary Report — Summary report allows you to store and later retrieve important ECG
and event information. You can print summary report information in various formats. For
more information about Summary Report and how to print a report, refer to the Summary
Report section below.

Full Disclosure Recording — Full Disclosure waveforms along with event information are
stored and may be reviewed using ZOLL CodeNet Central software. For information about
Full Disclosure Recording, refer to “Full Disclosure Recording” on page 10-8.

Incident Log — The Incident log is an abbreviated list of all major events recorded in
Summary Report. For more information about the Incident log and how to print it, refer to
“Printing an Incident Log” on page 10-8.

Summary Report

The R Series defibrillator automatically records defibrillation and cardioversion events,
PACER mode information, heart rate alarms, and segments of ECG when the recorder is
activated. Associated event information including device control settings, time, and date are
also recorded.

The following events trigger Summary Report to automatically record information:

Power is turned on.

Stripchart recorder is turned on.

Defibrillator shock is administered.

Code markers entered.

ECG rhythm analysis is initiated.

VF alarm is triggered.