ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 46
Advertising
R Series Service Manual
1-36
9650-0903-01 Rev. J
Note:
The calibration gas mixture and regulator are available from Respironics Novametrix.
• Gas Regulator: PN 6081-00
• Calibration Gas (carton of 4 tanks): PN 8964-00
21.10
Set the Gas Balance settings of the
CAPNOSTAT 5 CO
2
Sensor to that of the
calibration gas mixture (N
2
, N
2
O, or He).
The default gas balance is N
2
.
21.11
Allow a few seconds for the gas mixture to
stabilize and observe the CO2 Percent value.
The expected value is 5% ± 0.26%.
o
o
o
21.12
Press the Return softkey to return to the main
EtCO
2
Calibration screen.
21.13
Turn the device off when calibration is
complete.
Record your results on the Maintenance Test Checklist.
Do this...
Observe this...
Pass/Fail/NA