ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 46

Advertising
background image

R Series Service Manual

1-36

9650-0903-01 Rev. J

Note:

The calibration gas mixture and regulator are available from Respironics Novametrix.
• Gas Regulator: PN 6081-00
• Calibration Gas (carton of 4 tanks): PN 8964-00

21.10

Set the Gas Balance settings of the
CAPNOSTAT 5 CO

2

Sensor to that of the

calibration gas mixture (N

2

, N

2

O, or He).

The default gas balance is N

2

.

21.11

Allow a few seconds for the gas mixture to
stabilize and observe the CO2 Percent value.

The expected value is 5% ± 0.26%.

o

o

o

21.12

Press the Return softkey to return to the main
EtCO

2

Calibration screen.

21.13

Turn the device off when calibration is
complete.

Record your results on the Maintenance Test Checklist.

Do this...

Observe this...

Pass/Fail/NA