Defibrillator

Symptom

Recommended Action

1. Excessive artifact detected when

using paddles as ECG source.

Ensure “PADDLES” is selected.

Firmly press paddles against patient skin.

Use gel on paddles.

Clean paddle surface.

Check and clean between adult and pediatric shoe.

Check cable for damage.

Use ECG electrodes.

2. Defibrillator will not charge (energy

level does not increment on
display).

Ensure that SHOCK button(s) on paddles or front panel
are not stuck on.

Replace the battery pack.

3. Charge time to 200 J exceeds

10 seconds.

Typical in a low battery condition (up to 20 seconds)

Plug device into AC power.

Install fully charged battery pack.

4. Energy does not discharge when

the SHOCK button(s) is pressed.

60 seconds have elapsed in manual mode since initial
charge ready. Energy was internally discharged.

Device is in Sync mode or Remote Sync mode and no
QRS complex is detected.

Energy internally discharged because energy selection
was changed during charge or after the device was
ready.

Unit was not completely charged when SHOCK
button(s) were pressed. Wait for DEFIB XXXJ READY
message and ready tone.

Press and hold SHOCK button(s) until energy is
delivered to the patient.

Pads or paddles not making good contact with patient.

5. Unable to SHOCK when in Sync

mode

Ensure SYNC XXXJ SEL is displayed on monitor.

Make sure ECG signals are displayed.

Check for Sync markers (arrow above R wave). If not
present, change lead selection, or electrode placement.

Press and hold SHOCK button(s) until energy is
delivered to the patient.

Alter ECG electrode placement.