ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 40

R Series Service Manual
1-30
9650-0903-01 Rev. J
17.5
Increase the output to 80mA
Output on the Impulse 4000 is 80mA or +/- 5mA.
o
o
17.6
Increase the output to 120mA.
Output on the Impulse 4000 is 120mA or +/- 6mA.
o
o
17.7
Increase the output to 140mA.
Output on the Impulse 4000 is 140mA or +/- 7mA. Pulse width is 40mS +/-
2mS.
o
o
17.8
Decrease the output to 60mA.
Decrease the rate to 30 ppm.
Pacer rate on Impulse 4000 is 29-31 ppm.
o
o
17.9
Increase the rate to 180ppm.
Pacer rate on Impulse 4000 is 177-183 ppm.
o
o
17.10
Connect the ECG cable to the R Series and
Impulse 4000. Select the ECG at 60 BPM on
the Impulse 4000.
ECG at 60 BPM is seen on the display and no stimulus markers.
o
o
17.11
Press the Async Pace softkey.
ECG at 60 BPM seen on the display with the pace stimulus markers displayed.
Async pace message displays.
o
o
17.12
Turn off Impulse 4000. Set Pacer Rate to
100ppm. Press the RECORDER ON button.
Observe the pace stimulus markers every 15mm +/-1mm.
o
o
17.13
Press and hold 4:1 button.
Observe the pace stimulus markers every 60 mm+/- 1.5 mm.
o
o
Record your results on the Maintenance Test Checklist.
Do this...
Observe this...
Pass/Fail