18.0 SpO

2

Monitor Test (for SpO

2

Option)

Tools Needed:

Masimo



Reusable Sensor.

Masimo

Patient Cable.

Fluke Biomedical Index 2PFE SpO

2

Simulator (or equivalent).

Test Setup:

1. Connect the OneStep cable to the test port.

2.

DO NOT connect the ECG cable to the simulator.

3.

Install the Masimo

Patient Cable and attach the Masimo

sensor to the patient cable.

4.

Connect the Masimo

sensor to the finger simulation post.

5.

Place a fully charged battery into the battery well or connect to AC power.

6.

Ensure that the SpO

2

Simulator is off.

Do this...

Observe this...

Pass/Fail

18.1

Turn the selector switch to MONITOR(for AED
units turn to ON, and select MANUAL MODE.)

The SpO

2

saturation percentage appears as a dashed line on the monitor.

o

o

18.2

Wait ten seconds.

Turn on the SpO

2

simulator.

Press the SIM

softkey on the Index SpO

2

Simulator. Press the

MAN softkey.

The SpO

2

PULSE SEARCH message displays.

o

o

18.3

Press the 02+ or 02- softkey of the simulator
until the SpO

2

output is at 98%.

The R Series

SpO

2

reading of 98 +/- 1% appears on the R Series monitor.

Note that you may need to wait up to 2 minutes for the information to appear on
the ZOLL display.

o

o

18.4

Using the Index

SpO

2

Simulator, press the

BPM+ or BPM- softkey until the heart rate is
230 BPM.

The

SpO

2

rate 230 BPM displays on the simulator screen.

Note that you may need to wait up to 2 minutes for the information to appear on
the ZOLL display.

The

SpO

2

saturation of 96-100% appears on the R Series display.

The heart

rate of 226-234 BPM displays on the R Series monitor.

o