0 front panel button test, Front panel button test -6 – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 16

Advertising
background image

R Series Service Manual

1-6

9650-0903-01 Rev. J

2.0 Front Panel Button Test

Tools Needed: Fluke Impluse 4000.

Test Setup:

1.

Install strip chart paper into the printer compartment.

2.

Install the battery in the unit or connect the A/C power cord to the unit and then plug the cord into an electrical outlet.

3.

Connect the OneStep

TM

cable and ECG cable (3 lead adapter (part #8009-0762-XX), 3 lead cable, or 5 lead cable) to the Fluke Biomedical 4000 Analyzer

(or equivalent).

Do this...

Observe this...

Pass/Fail/NA

2.1

Turn the selector switch to MONITOR. (for AED
units, turn the selector switch to ON and select
Manual mode.)

Listen for 4 beep tones. PADS and MONITOR display on the monitor.

NOTE: PADS is a factory default setting.

o o o

2.2

Press the LEAD button; three times for the 3
lead cable and seven times for the 5 lead cable.

Each time you press the LEAD button, a different lead number appears under
the LEAD heading on the display.

PADS, I, II, III will display if a 3 lead ECG cable is connected or no ECG cable
is connected.

PADS, I, II, III, AVR, AVL, AVF, V1 will display if a 5 lead ECG cable is
connected.

o o o

2.3

Set the Impulse 4000 to NSR of 120 BPM.
Press the LEAD button until Lead II displays. To
check the size of the ECG waveform, press the
SIZE button.

As you press the SIZE button five times (0.5, 1.0, 1.5, 2.0, 3.0), note that the
size of the ECG waveform appropriately changes on the display.

o o o

2.4

Press the ALARM SUSPEND button.

Bell changes from disabled to enabled. If the alarm sounds, press the ALARM
SUSPEND button to turn it off. The alarm will only be suspended for 90
seconds at this point. Press and hold the ALARM SUSPEND button for 3
seconds to disable alarms.

o

o

o

2.5

Press the RECORDER button.

The strip chart paper moves out of the unit from the printer compartment.
Check that the correct time, date, ECG lead annotation and waveform are
recorded on the paper. (Set Time and Date, if necessary.)

o

o

o

2.6

Open the printer door.

Press RECORDER button.

CHECK RECORDER message appears on the monitor.

o

o

o