R Series Service Manual

9650-0903-01 Rev. J

1-

11

Test Setup:

1. Remove red lead from power supply and connect to 0.1



resistor

.

2.

Connect other end of resistor to “+” terminal of power supply using a second red lead.

3.

Connect multimeter across the resistor.

4.

Set voltage scale (if DVM is not autoranging) to 220 mV.

Do this...

Observe this...

Pass/Fail/NA

4.7

System Current Test

Set power supply to 10.8V.

4.8

Turn the selector switch to MONITOR.

Voltage across resistor should be 145 mV or less (<1.2A of ON current).
NOTE: Without optional parameters.

All devices with SpO

2

, EtCO

2

or NIBP <160mV

o

o

o

o

o

o

4.9

Turn unit off.

Record your results on the Maintenance Test Checklist.

15 Amp