Defib 120j sel, 2 press analyze button – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 419

Advertising
background image

Advisory Defibrillation Procedure

9650-0912-01 Rev. K

ZOLL R Series Operator’s Guide

4–3

When used with OneStep Pediatric electrodes, the default energy selections for pediatric
patients are:

Shock 1 - 50 joules

Shock 2 - 70 joules

Shock 3 - 85 joules

WARNING!

Use only OneStep Pediatric electrodes to defibrillate patients under 8 years of age in
Advisory mode. Use of adult electrodes, or pediatric electrodes other than OneStep
Pediatric electrodes, can result in the delivery of excessive energy doses.

After the third shock, all subsequent shocks are delivered at the same energy as the third shock
in both Adult and Pediatric modes.

If medical protocols allow, you may select a different energy level using the energy select up
and down arrow buttons on the front panel. The new energy setting displays on the monitor.

If you have configured SHOCK 1, 2, and 3 to escalating energy levels, and then you manually
change the energy level outside preconfigured SHOCK 1, 2, 3 sequence and deliver a shock, it
disables the automatic energy escalation. See the Energy Level: Shock 1, 2, 3 section of the
R Series Configuration Guide for more details.

2 Press ANALYZE Button

WARNING!

Keep patient motionless during ECG analysis. Do not touch the patient during analysis.
Cease all movement via stretcher or vehicle before analyzing the ECG.

DEFIB 120J SEL.

Code

Marker

Options

00:01

ECG

Param

Report

Data

Sync

On/Off

Alarms

Advertising