ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 221

Advertising
background image

AED Operation

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

3–9

SHOCK ADVISED/ XXXJ READY

This message is displayed when a shockable ECG rhythm has been detected and the selected
energy is ready to be delivered.

PRESS SHOCK

This message is displayed and announced when the ECG analysis has determined that a shock
is advised and the selected energy is ready to be delivered.

SHOCKS: XX

This message displays the number of shocks that have been delivered by the unit since it was
powered on. It resets to 0 after the unit has been off for more than 10 seconds.

NO SHOCK ADVISED

When ECG analysis detects a nonshockable rhythm, this message displays and continues for 10
seconds following completion of the analysis.

CHECK PULSE

If configured to do so, this message is displayed and announced in the following situations:

After a No Shock Advised analysis result

During the CPR interval after a No Shock Advised analysis result

During the CPR interval after delivery of the final shock in a stacked shock sequence

After the final shock is delivered

IF NO PULSE, PERFORM CPR

If configured to do so, this message is displayed and announced in the following situations:

During the CPR interval after a No Shock Advised analysis result

During the CPR interval after delivery of the final shock in a stacked shock sequence

During the start of extra CPR interval

PERFORM CPR

If configured to do so, this message is displayed and announced in the following situations:

During the CPR interval after a No Shock Advised analysis result

During the CPR interval after delivery of the final shock in a stacked shock sequence

PUSH HARDER

This message is announced when the chest compressions applied during CPR are not forceful
enough.

GOOD COMPRESSIONS

This message is announced when the chest compressions applied during CPR are forceful
enough.

CHECK PADS

This message is displayed and announced when the therapy pads have been disconnected from
the patient.

Advertising