R Series Service Manual

9650-0903-01 Rev. J

3-9

5. Remove the 3 screws securing the battery well.

6. Lift the battery well from the front of the R Series, tilting

towards the rear. DO NOT disconnect the low and high
current harnesses. Feed them up through the chassis to
remove the battery well.

Note:

During reassembly, torque to 6 in-lbs.