Sync 120j sel, 2 charge defibrillator – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 428

Advertising
background image

C

HAPTER

5

S

YNCHRONIZED

C

ARDIOVERSION

5–4

www.zoll.com

9650-0912-01 Rev. K

A SYNC XXXJ SEL. message appears on the display. If DEFIB XXXJ SEL. appears, press the
Sync On/Off softkey. (If your unit supports Remote Sync, you must also press the Sync
softkey.) Two quick beeps sound.

Unless otherwise configured, the unit automatically exits Sync mode after each shock and when
the Mode Selector is moved to MONITOR, PACER or OFF.

To reactivate Sync mode, press the Sync On/Off softkey again. (If your unit supports Remote
Sync, press the Sync softkey again.) Changing the selected energy levels does not cause the
unit to leave Sync mode.

Note that the unit can be configured to stay in Sync mode after defibrillation, if desired. Refer
to the R Series Configuration Guide for instructions.

2 Charge Defibrillator

Press the CHARGE button on the front panel or on the apex paddle handle.

To abort charging and increase or decrease the selected energy after the CHARGE button has
been pressed, use the ENERGY SELECT buttons on either the defibrillator front panel or the
sternum paddle. Press the CHARGE button again to charge the unit to the newly selected
energy level.

After charging the unit to the selected energy, either the front panel SHOCK button or the
APEX paddle charge indicator illuminates. A distinctive audible tone sounds and the SYNC
XXXJ READY
message is displayed.

The defibrillator is now ready to deliver therapy.

SYNC 120J SEL.

Code

Marker

Options

00:01

ECG

Param

Report

Data

Sync

On/Off

Alarms

3

SHOCK

SHOCK

ENERGY

SELECT

ENERGY

SELECT

ANALYZE

ANALYZE

CHARGE

CHARGE

2

1

RELEASE

RELEASE

or

Advertising