ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 52

Advertising
background image

R Series Service Manual

1-42

9650-0903-01 Rev. J

5

If you are using the Fluke® Biomedical CuftLink Simulator, the volume leak reading should be <10.

6

If you are using the Fluke® Biomedical CuftLink Simulator, the volume leak reading for a failure should be >10.

24.5

On the NIBP simulator, press the START TEST
softkey.

Note: You must press the START TEST

softkey within 30 seconds of closing the
valves on the R Series unit.

After approximately 1 minute, a number appears in the upper middle area of
the NIBP simulator display.

If the simulator:

displays a Volume Leak reading <5

5

, then the R Series unit has passed the

test.

displays a Volume Leak reading >5

6

, then the R Series unit has failed the

test.

displays no Volume Leak reading, but maintains a stable pressure reading
at or above 200 mmHg, then the R Series unit has passed the test; there is
no volume leak.

In addition, the R Series displays the simulator’s pressure reading in the “Cuff
Pressure” field.

After approximately 3 minutes, the valves on the R Series unit open.

o

o

24.6

On the NIBP simulator, press the STOP TEST
softkey.

The NIBP simulator terminates the Volume Leak Test.

o

o

24.7

On the R Series unit, press the EXIT softkey
twice.

The R Series returns to the main NIBP Service Mode screen, then to normal
Monitor mode operation.

Record your results on the Maintenance Test Checklist.

Do this...

Observe this...

Pass/Fail