Discharging capacitor, Capacitor discharging -28 – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 94

Advertising
background image

R Series Service Manual

3-28

9650-0903-01 Rev. J

Discharging Capacitor

WARNING! This unit may contain lethal voltages. You MUST completely discharge the high voltage capacitor before removing from unit. DO NOT SHORT

THE TERMINAL ENDS OF THE CAPACITOR.

1. While capacitor is connected to the PD Engine

assembly, discharge the capacitor by bleeding out
the excess voltage using a resistor with values of
approximately 5 kohms, 25 watts for 10-20 seconds.
Apply the resistor across the terminals on the board.
Measure the voltage on the capacitor terminals to
verify that capacitor is discharged.