R Series Service Manual

9650-0903-01 Rev. J

1-

27

15.0 Summary Report Test

Tools Needed: Impulse 4000.

Test Setup:

1. Connect the OneStep cable to the Impulse 4000

2.

If you are using paddles, place the paddles on the analyzer’s discharge plates.

Do this...

Observe this...

Pass/Fail

15.1

Press the Report Data * softkey, press Erase,
then Erase All.

ERASING REPORT displays. Wait for the message to clear.

o

o

15.2

Set selector switch to DEFIB(for AED units turn
to ON, and select MANAUL MODE.) Select
200J using the ENERGY SELECT button, and
press the CHARGE button. When charged,
press the SHOCK button to discharge into the
defibrillator analyzer.

The unit successfully discharges and if configured to do so,prints a strip chart.

o

o

15.3

Wait 18 seconds, then press the Code Marker
softkey. Press the CPR softkey.

The Code Markers display. (After CPR Code Marker is selected, the unit
automatically exits Code Marker options menu.)

o

o

15.4

Turn the unit off. Wait 10 seconds and then turn
the unit on. Press the Report Data * softkey,
then press Print Chart, then Print All.

Summary report prints. The report displays the correct date, time, the shock
delivered and Code Marker event.

o

o

* For software versions 3.0 or earlier, the unit displays Report instead of Report Data.

Record your results on the Maintenance Test Checklist.