R Series Service Manual

1-12

9650-0903-01 Rev. J

Test Setup for Off Current Test:

1. Remove 0.1

resistor and replace with 1K

.

2.

Connect DMM across resistor.

3.

Set voltage scale to DCV.

4.

Measure voltage across resistor.

Charger Test

Tools Needed: 2-Post Battery Fixture (9100-0575-TF), DVM, 2 Test Leads, Impulse 4000, Stopwatch.

Test Setup:

Set DVM to read DC Volts.

Connect Postive lead to positive post of the R Series charger load fixture.

Connect negative lead to negative post of the R Series charger load fixture.

Verify the unit is plugged into AC Power.

Do this...

Observe this...

Pass/Fail

4.10

Off Current Test

Measure across resistor with unit turned off.

Voltage should be less than 270 mV (<270

A of current).

o

o

Record your results on the Maintenance Test Checklist.

Do this...

Observe this...

Pass/Fail

4.11

Set front panel switch to DEFIB or ON, (for BLS
units, press MANUAL MODE, then press
CONFIRM).

4.12

Install Charger Test Fixture.