0 advisory message test (manual/advisory units), Advisory message test (manual/advisory units) -28 – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 38

Advertising
background image

R Series Service Manual

1-28

9650-0903-01 Rev. J

16.0 Advisory Message Test (Manual/Advisory Units)

Tools Needed: Impulse 4000

Test Setup:

1. Connect the OneStep cable via the adapter (D.N.I #3010-0378) to the defibrillator analyzer.

Do this...

Observe this...

Pass/Fail

16.1

Connect OneStep cable to the simulator.

Turn the selector switch to DEFIB mode (for
AED units turn to ON, and select MANUAL
MODE.)

16.2

Select VF (ventricular fibrillation) on the Impulse
4000, then press the ANALYZE button.

ANALYZING ECG message displays.

STAND CLEAR message displays.*

SHOCK ADVISED message displays.

PRESS SHOCK message displays*+

*Advisory audio prompts are user configurable.

+If configured for auto charge. PRESS CHARGE message displays if not
configured to auto charge.

o

o

16.3

Press the SHOCK button.

Unit discharges.

o

o

16.4

Select the NSR (normal sinus rhythm) on the
simulator, then press the ANALYZE button.

ANALYZING ECG message.

STAND CLEAR message.*

NO SHOCK ADVISED message.*

*Advisory audio prompts are user configurable.

o

o

Record your results on the Maintenance Test Checklist.