Intended use — spo, Monitoring, Intended use — etco – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 380: Pediatric pacing

Advertising
background image

C

HAPTER

1

G

ENERAL

I

NFORMATION

1–10

www.zoll.com

9650-0912-01 Rev. K

Pediatric Pacing

Pacing can be performed on pediatric patients weighing 55 lb. (25 kg) or less using ZOLL
pediatric hands-free therapy electrodes. Prolonged pacing (in excess of 30 minutes),
particularly in neonates, can cause burns. Periodic inspection of the underlying skin is
recommended.

Intended Use — SpO

2

Monitoring

The R Series pulse oximeter, with the Masimo

®

SET

®

technology and the LNCS

®

series of

oximeter sensors, is indicated for the continuous, noninvasive monitoring of arterial oxygen
saturation (SpO

2

) and pulse rate during both no motion and patient motion conditions for adult

patients, and no motion conditions for pediatric and neonatal patients in a hospital or
prehospital environment.

Intended Use — EtCO

2

Monitoring

The ZOLL R Series EtCO

2

option with Respironics Novametrix technology is indicated for the

continuous noninvasive monitoring of end tidal carbon dioxide (EtCO

2

) and respiration rate in

patients requiring ventilator support, in-hospital transport, or anesthesia.

This option uses the CAPNOSTAT 5 Mainstream CO

2

sensor attached to an airway adapter that

connects to an endotracheal tube, mask or disposable mouthpiece.

The R Series EtCO

2

option is designed to monitor adult, pediatric, and neonatal patients.

The following substances can influence CO

2

measurements made with the CAPNOSTAT 5 CO

2

sensor:

elevated oxygen levels

nitrous oxide

halogenated agents

The R Series EtCO

2

option provides settings for high oxygen and/or nitrous oxide

compensation. Halogenated anesthetic agents alter CO

2

readings, but the R Series unit will

monitor CO

2

within specifications when these agents are present at normal clinical levels. The

presence of Desflurane in the exhaled breath beyond normal values (5%) may positively bias
measured carbon dioxide values by up to an additional 3 mmHg.

The R Series EtCO

2

option is intended for use only with the ZOLL/Respironics Novametrix

CAPNOSTAT 5 Mainstream CO

2

Sensor and mainstream airway adapters.

The R Series EtCO

2

option can be used on adult patients (21 years of age and older) and on

pediatric patients, as described in the following table:

Pediatric Subpopulation

Approximate Age Range

Newborn (neonate)

Birth to 1 month of age

Infant

1 month to 2 years of age

Child

2 to 12 years of age

Adolescent

12-21 years of age