Heart rate alarm limits, Alarms – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 459

Advertising
background image

Alarms

9650-0912-01 Rev. K

ZOLL R Series Operator’s Guide

9–9

To set alarm parameters:

1. Press the Alarms softkey to view the Alarm Set screen and softkeys.

2.

Press the Next Param or Prev Param softkey.

This scrolls the highlighted area among the different available vital signs.

If you want to change the state of the highlighted vital sign:

a. Press the Change Value softkey.

b. Press the Inc > or Dec < softkey to change the state value.

c. Press the Enter softkey.

The State field can be set to three possible values, Enable, Disable, or Auto.

Disable permanently turns off alarm processing for the selected physiological parameter.

Enable causes alarm processing to operate whenever alarms are activated via the front
panel ALARM key.

Selecting AUTO sets the lower and upper alarm limits to 80% and 120% of the patient’s
currently measured heart rate, if valid measurements are present for the vital sign when
the Enter softkey is pressed. (Refer to appropriate Operator’s Guide parameter insert(s)
for percentages associated with other parameters).

3.

Press the Next Field softkey to move to the Low or High field for the highlighted vital sign;
repeat steps 2a through 2c to change the Low or High value.

Note:

To recalculate the Low and High limits for any parameter when these limits have
previously been set using the AUTO State, follow the procedure above to select AUTO
again, and then press the Enter softkey. The unit automatically resets the Low and
High limits based upon the currently measured value of the selected physiological
parameter.

4.

Press the Return softkey to set all values and return to normal operating mode.

Heart Rate Alarm Limits

The heart rate is displayed in the upper right-hand corner of the screen.

Unless configured otherwise, heart rate alarms are preset at 30 bpm (bradycardia) and 150 bpm
(tachycardia). The low heart rate alarm limit range is 20 bpm to 100 bpm.

Prev

Param

Next

Field

Return

72

ECG

PADS

x1

SpO2%

0:00

IDLE

Depth
Rate

99

NIBP

mmHg

---
---

---

CO2

mmHg

---

---

RR

Ɔ

Parameter

State

Low

High

SpO2

ENABLE

DISABLE

30

85

150
100

ECG HR

ALARM SET

Next

Param

Change

Value

EtCO2

RESP RATE

NIBP SYS

NIBP DIA

NIBP MEAN

ENABLE

ENABLE

ENABLE

ENABLE
ENABLE

25

55

5

120

90

160

50

110

60

130

CPR

PPI

Each setting includes the alarm state
(ENABLE, DISABLE, or AUTO) and
the low and high limit of the
acceptable range of values.