R Series Service Manual

2-4

9650-0903-01 Rev. J

CHECK RECORDER

Produced when paper tray is empty, paper jams or
recorder door is opened.

CHECK SPO2 SENSOR

Reposition SpO

2

sensor on patient.

CLOCK BATTERY FAULT

The RTC coin battery has failed.

Replace lithium coin battery or digital board.

CLOCK FAULT 11

Real time clock oscillator failure

Replace lithium coin battery or digital board.

CLOCK FAULT 12

Real time clock back-up power supply failure. Found
oscillator stopped at power-up, but oscillator now running
when the system is running (oscillator only runs when
main power is applied).

Replace lithium coin battery or digital board.

CLOCK FAULT 13

One of the set time units (seconds, minutes, year, etc.) is
out of range.

Replace lithium coin battery or digital board.

CO2 COMM ERROR

No or invalid communication from the EtCO

2

module.

Replace EtCO

2

module and or system board.

CO2 DEVICE NOT READY

There is CO

2

in the airway adapter when attempting to

zero. Zeroing was attempted within 20 seconds of
previous zero operation.

Remove airway adapter from CO

2

source

including the patient’s, and your own exhaled
breaths, and ventilator exhaust valves.

Wait up to 20 seconds before retrying a
mainstream airway adapter zero, as described
in “Zeroing the Mainstream CAPNOSTAT 5
CO

2

Sensor/Airway Adapter”

CO2 IN LINE: WAIT

Adapter zero attempted with CO

2

in the adapter.

CO2 MODULE NOT VALID

Sidestream sensor connected (not supported by the
R Series unit’s operating software)

Use mainstream sensor.

CO2 OUT OF RANGE

The calculated CO

2

value is greater than 150 mmHg.

If error persists, perform a mainstream airway
adapter zero, as described in “Zeroing the
Mainstream CAPNOSTAT 5 CO

2

Sensor/Airway

Adapter.”

Error Message

Explanation

User
Advisory

Technical Action