R Series Service Manual

1-32

9650-0903-01 Rev. J

18.5

Using the Index

SpO

2

Simulator, press the

BPM- softkey until the heart rate is 50 BPM.

The

SpO

2

saturation of 96-100% displays on the unit.

The heart rate of 46-54 BPM displays on the R Series monitor.

o

o

o

o

18.6

Using the Index

SpO

2

Simulator, press the 02+

softkey until the SpO

2

output is at 72%.

The

SpO

2

saturation of 69-74% displays on the unit.

The heart rate of 46-54 BPM displays on the R Series monitor.

o

o

o

o

18.7

Press the OPTIONS softkey. Press TRACES.

Press TRACE 2. Select

SpO

2

.

Plethysmographic waveform appears on the ZOLL display.

o

o

18.8

Press RECORDER.

The plethysmographic waveform prints on the strip chart paper.

o

o

18.9

Press RECORDER to stop printing.

Verify that the recorder stops.

o

o

18.10

Using the Index

SpO

2

Simulator, press the

BPM- softkey until the heart rate is at 230 BPM.

The

SpO

2

saturation rate of 69-74% displays on the unit.

The heart rate in the heart position of 226-234 BPM displays on the monitor.

o

o

18.11

Press RECORDER.

Verify that the waveform is printed at the correct rate.

Press RECORDER to stop printing.

o

o

18.12

Press RECORDER. Remove the Masimo

patient cable.

Verify that the recorder stops.

o

o

Record your results on the Maintenance Test Checklist.

Do this...

Observe this...

Pass/Fail