ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 57

Advertising
background image

R Series Service Manual

9650-0903-01 Rev. J

2-5

CO2 UNIT ERROR

The EtCO

2

sensor or module has detected a hardware

error.

Check that the sensor is properly plugged in.
Re-insert the sensor. Turn R Series unit off, then
on again to reset. perform a mainstream airway
adapter or module zero, as described in “Zeroing
the Mainstream CAPNOSTAT 5 CO

2

Sensor/

Airway Adapter”.

If the problem persists, contact ZOLL Technical
Support.

CO2 WARM UP

The mainstream sensor is warming up. This may take up
to 5 minutes.

Wait for sensor or module to warm up.

If the message persists more than 5 minutes,
replace the sensor.

CPR FAULT 8

ECG processor not receiving CPR data

Replace analog board.

DATA TRANSFERRED

Transfer done message

DEFIB DISABLED

User prompt issued simultaneously with other faults if defib
is disabled.

Possible configuration problem. Replace high
voltage module. Call ZOLL Technical Support.

DEFIB FAULT 76

PD Defib failure during POST.

Replace PD engine.

DEFIB FAULT 77

PD Defib failure while running.

Replace PD engine.

DEFIB FAULT 78

PD Defib functional safety error while running.

Replace PD engine.

DEFIB FAULT 79

PD module Defib/Pace failure during POST.

Replace PD engine.

DEFIB FAULT 80

Undefined error received from the PD module.

Replace PD engine.

DEFIB FAULT 94

PD module reset 3 times without being requested to.

Replace PD engine.

DEFIB FAULT 95

PD module not communicating.

Replace PD engine.

DEFIB FAULT 96

PD module reported an error on the discharge.

Replace PD engine.

DEFIB MAINT. REQ.

More than 5000 discharges of 200J have occurred.
Maintenance is required.

Replace PD engine.

Error Message

Explanation

User
Advisory

Technical Action