23.0 NIBP Monitor Test

The NIBP monitor test verifies the repeatability of the systolic, diastolic, and mean blood pressure measurements, as well as the patient pulse rate
calculation.

Tools Needed: NIBP simulator (The values and procedure provided in this manual are specific to the BP Pump 2)

Note

The primary propose of an NIBP simulator is to reproduce a pressure profile similar to a live patient to be used for testing repeatability and
functionality of the system. There are many different NIBP simulators on the market, each manufacturer uses a different method to develop their
algorithm, and as such this can cause readings from different simulators to vary. In order to test for repeatability, you should first establish the
offset

1

of your simulator. The offset value should then be used to determine the expected values. NIBP simulators cannot be used as a source for

testing the accuracy of the non-invasive blood pressure measurements of monitors such as the ZOLL R Series.

Test Setup:

1. Connect the simulator hose to the NIBP connector on the R Series unit.

2.

Set the following parameters on the NIBP simulator

2

:

3.

Make sure the ECG cable is not connected to the R Series unit.

4.

If the SpO

2

option is installed, make sure that the SpO

2

patient cable is NOT connected to the R Series unit.

1

NIBP Simulators may produce a reading on the NIBP monitor that is shifted from the simulator's setting. The offset value must be established based on a

statistical sample of monitors and readings. Please contact ZOLL Technical Support if you require assistance establishing the offset of the simulator and test
set-up that you are utilizing.

2

If you are using the Fluke

®

Biomedical CuftLink, you must change the shift value of the Blood Pressure Envelope to +3 on the Pressure Curve Adjust Menu.

Parameter

Value

Systolic pressure

120 mmHg

Diastolic pressure

80 mmHg

Mean pressure

93 mmHg

3

Heart rate

80 bpm