Manual Defibrillator Testing
9650-0912-01 Rev. K
ZOLL R Series Operator’s Guide
12–7
Figure 12-1.
Code Readiness Test Log Print Display
Figure 12-2.
Code Readiness Test Log
Figure 12-3.
Code Readiness Test Report
Code Readiness Test Log
Test
Return
Prev
Test
Next
Test
Date: 10 NOV 06 00:01:03
Type: Automatic
Defib: OK
Pace: OK
ECG: OK
AC Power: Connected
Battery: OK
30J Self Test: OK
OneStep Cable: OK
Test Log