ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 58

R Series Service Manual
2-6
9650-0903-01 Rev. J
DEFIB NOT CHARGED
Discharge button is pressed in a Defib mode but the unit is
not charged.
DEFIB OVERUSE
More than 50 shocks were delivered in less than 20
minutes.
Unit needs to cool down, wait approximately
5 minutes.
DEFIB PAD SHORT
Measured impedance between high voltage leads of MFC.
Ensure pads are coupled to patient. Check/
replace pads or universal cable. Replace system
board.
DISK FORMAT REQ.
Report error if any problem with DOC file access occurred.
Replace digital board.
ECG DISABLED
Persistent Critical Hardware Failure on ECG module.
Replace analog board.
ECG FAULT 200
No ECG applicable available.
Reload MCU Software and ECG App. Software.
ECG FAULT 3
ECG processor reset failure.
Replace analog board.
ECG FAULT 4
Excessive number of missed samples from ECG
processor.
Replace analog board.
ECG FAULT 5
ECG processor power up failure.
Replace analog board.
ECG FAULT 6
ECG processor Hardware failure.
Replace analog board.
ECG FAULT 7
ECG processor is not responding.
Replace analog board.
ECG LEAD OFF
1 or more ECG leads are not connected when non-MFE
leads are selected as input.
ECG RA LEAD OFF
ECG lead RA is disconnected
ECG TOO LARGE
Number of intervals with large ECG exceeds threshold.
Issued immediately before RETRY ANALYSIS.
ECG V LEAD OFF
ECG lead V is disconnected
ENABLE ETCO2
User attempts to zero EtCO
2
sensor while the EtCO
2
feature is disabled
Using the softkeys, select Enable EtCO2.
Error Message
Explanation
User
Advisory
Technical Action