ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 58

Advertising
background image

R Series Service Manual

2-6

9650-0903-01 Rev. J

DEFIB NOT CHARGED

Discharge button is pressed in a Defib mode but the unit is
not charged.

DEFIB OVERUSE

More than 50 shocks were delivered in less than 20
minutes.

Unit needs to cool down, wait approximately
5 minutes.

DEFIB PAD SHORT

Measured impedance between high voltage leads of MFC.

Ensure pads are coupled to patient. Check/
replace pads or universal cable. Replace system
board.

DISK FORMAT REQ.

Report error if any problem with DOC file access occurred.

Replace digital board.

ECG DISABLED

Persistent Critical Hardware Failure on ECG module.

Replace analog board.

ECG FAULT 200

No ECG applicable available.

Reload MCU Software and ECG App. Software.

ECG FAULT 3

ECG processor reset failure.

Replace analog board.

ECG FAULT 4

Excessive number of missed samples from ECG
processor.

Replace analog board.

ECG FAULT 5

ECG processor power up failure.

Replace analog board.

ECG FAULT 6

ECG processor Hardware failure.

Replace analog board.

ECG FAULT 7

ECG processor is not responding.

Replace analog board.

ECG LEAD OFF

1 or more ECG leads are not connected when non-MFE
leads are selected as input.

ECG RA LEAD OFF

ECG lead RA is disconnected

ECG TOO LARGE

Number of intervals with large ECG exceeds threshold.
Issued immediately before RETRY ANALYSIS.

ECG V LEAD OFF

ECG lead V is disconnected

ENABLE ETCO2

User attempts to zero EtCO

2

sensor while the EtCO

2

feature is disabled

Using the softkeys, select Enable EtCO2.

Error Message

Explanation

User
Advisory

Technical Action