5 cal meas ph, 1 cal meas ph - overview, Cal meas ph – Metrohm tiamo 2.2 Manual User Manual
Page 1048

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1034
■■■■■■■■
tiamo 2.2
Temp. (°C)
Pre4
Pre7
Pre9
pH 4.00
pH 7.00
pH 9.00
30
4.00
6.99
8.96
35
4.01
6.98
8.93
40
4.02
6.98
8.90
45
4.03
6.97
8.87
50
4.04
6.97
8.84
55
4.06
6.97
8.81
60
4.07
6.97
8.79
65
4.09
6.98
8.76
70
4.11
6.98
8.74
75
4.13
6.99
8.73
80
4.15
7.00
8.71
85
4.18
7.00
8.70
90
4.20
7.01
8.68
95
4.23
7.02
8.67
5.6.5.5
CAL MEAS pH
5.6.5.5.1
CAL MEAS pH - Overview
Dialog window: Method
▶ CAL MEAS pH ▶ Properties.. ▶ CAL MEAS pH - 'Command
name'
Measuring command for Calibration with pH electrodes, which is used
together with the command CAL LOOP pH.
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 841, 842, 857, 888, 901, 902, 904, 905,
906, 907
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798, 799
pH/Ion - Meter: 867
Robotic Titrosampler: 855
Appearance
The command has the following appearance: