5 Method

tiamo 2.2

■■■■■■■■

649

or after a waiting period. The volume that has been added when the end-
point is reached gives the calculable reagent consumption.

Reagent addition takes place in three phases during the titration:

■

Initial dosing
During this phase the dosing rate increases continuously. It starts with
the Min. rate and increases to the Max. rate.

■

Continuous dosing
In this phase addition is carried out at the Max. rate until the Control
range
is reached.

■

Dynamics
Dosing is finely controlled in this range. Shortly before the endpoint is
reached addition is only carried out at the Min. rate (Control range).