5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
968
■■■■■■■■
tiamo 2.2
Stop meas. value
Stops when the preset measured value has been reached since the start of
the measurement.
Range
0.0000 ... 2000.00 mS/cm
Selection
off
Default value
off
off
No stop.
Temperature
Measurement temperature
Temperature of the solution to be measured. If a temperature sensor is con-
nected, the temperature entered here is ignored.
712
Range
-170.0 ... 500.0°C
Default value
20.0
856
Range
-20.0 ... 150.0°C
Default value
20.0
Reference temperature
The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically conver-
ted to this reference temperature (usually 20 °C or 25 °C).
712
Range
-170.0 ... 500.0°C
Default value
20.0
856
Range
-20.0 ... 150.0°C
Default value
20.0
Selection
off
off
The conductivity is displayed for the measurement temperature.
Temperature compensation
Selection
Temperature coefficient | Sample solution
Default value
Temperature coefficient
Temperature coefficient
Manual input of a constant temperature coefficient.