Metrohm tiamo 2.2 Manual User Manual
Page 1430

7.12 855 Robotic Titrosampler
■■■■■■■■■■■■■■■■■■■■■■
1416
■■■■■■■■
tiamo 2.2
Lift rate
Lift speed for manual operation.
Range
3 ... 25 mm/s
Default value
25 mm/s
Axial distance
Distance between the axis of rotation of the sample rack and swing axis of
the robotic arm.
Range
100.0 ... 300.0 mm
814 USB Sample Processor XL
Selection
166.0 mm
Default value
166.0 mm
815 Robotic USB Sample Processor XL, 855 Robotic Titrosampler, 864 Bal-
ance Sample Processor
Selection
196.0 mm
Default value
196.0 mm
Swing Head
This parameter is only shown when a swing head is mounted on the tower.
Serial number
Shows the serial number of the swing head.
Swing position
Lift position valid for all 4 external positions, at which the swing head turns
to the external positions.
Range
0 ... 235 mm
Default value
0 mm
Rinse position
Rinse position valid for all 4 external positions.
Range
0 ... 235 mm
Default value
0 mm
[Configuration]
Opens the dialog window Configuration (see Chapter 7.11.7, page
1411) in which the properties of the robotic arm can be editet.