3 request - sample data request – Metrohm tiamo 2.2 Manual User Manual
Page 1196

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1182
■■■■■■■■
tiamo 2.2
5.6.10.2.3
REQUEST - Sample data request
Dialog window: Workplace
▶ [START] ▶ Sample data request
Message
Displaying the message defined in the command REQUEST.
Sample position
Position of the sample on the Sample Processor rack. This number can be
used for moving to a sample position with the command MOVE when
Target position = Sample position.
Range
1 ... 999
Default value
1
ID1 … ID16
Sample identifications.
Type = Number
Range
-1.0E-99 ... 1.0E+99
Type = Text
Input
max. 100 characters
Type = Date/Time
Selection
'YYYY-MM-DD' or 'YYYY-MM-DD hh:mm:ss'
Note
If the sample identifications are still invalid at the start of the method and
are not entered until the time of the REQUEST command during the run,
then the option Check at start in the START command must be
switched off for these method variables.
Sample size
Sample size (initial weight).
Input
1 ... 10 digits.
Default value
1
Device
Unit of sample size.
Selection
g | mg | µg | mL | µL | pcs
Default value
g