Metrohm tiamo 2.2 Manual User Manual

Page 865

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2

■■■■■■■■

851

Titrando, 855
Range

0.1 ... 999999 s

Default value

2.0 s

Titrino
Range

1 ... 999999 s

Default value

2 s

Start time

The measured values are not entered in the measuring point list until this
time has elapsed since the start of the titration.

Range

0 ... 999999 s

Default value

0 s

Start measured value

The measured values are not entered in the measuring point list until this
value has been reached.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Selection

off

Default value

off

Titrino
Range

-2000 ... 2000 mV

Selection

off

Default value

off

Start rate

The measured values are not entered in the measuring point list until the
dosing rate has fallen below this value. This parameter is checked 10 s after
the start at the earliest.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

off

Default value

off

Titrino
Range

0.01 ... 150.00 mL/min

Selection

off

Default value

off