Metrohm tiamo 2.2 Manual User Manual
Page 865

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
851
Titrando, 855
Range
0.1 ... 999999 s
Default value
2.0 s
Titrino
Range
1 ... 999999 s
Default value
2 s
Start time
The measured values are not entered in the measuring point list until this
time has elapsed since the start of the titration.
Range
0 ... 999999 s
Default value
0 s
Start measured value
The measured values are not entered in the measuring point list until this
value has been reached.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Selection
off
Default value
off
Titrino
Range
-2000 ... 2000 mV
Selection
off
Default value
off
Start rate
The measured values are not entered in the measuring point list until the
dosing rate has fallen below this value. This parameter is checked 10 s after
the start at the earliest.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
off
Default value
off
Titrino
Range
0.01 ... 150.00 mL/min
Selection
off
Default value
off