5 Method
tiamo 2.2
■■■■■■■■
853
Last dosing
The stop time starts after the last increment has been dosed, i. e. the stop
time is reset to zero again after each dosing step.
Stop rate
Rate at which the dosing cylinder is to be filled after titration. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
off
Default value
off
Titrino
Range
0.01 ... 150.00 mL/min
Selection
off
Default value
off
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration. The
maximum filling rate depends on the cylinder volume of the exchange unit
or dosing device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.8.7.7
STAT U - Monitoring
Tab: Method
▶ STAT U ▶ Properties... ▶ Monitoring
Command name
Name of the command.
Input
25 characters
The following monitoring processes can be switched on and defined on this
tab: