4 cal cond - conductivity standard – Metrohm tiamo 2.2 Manual User Manual
Page 1068

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1054
■■■■■■■■
tiamo 2.2
856
Range
0.1 ... 999999.0 s
Default value
2.0 s
Measurement without drift control
If the option Measurement without drift control has been selected, then
the following parameters will be displayed:
Measuring time
Maximum period of time for measurement.
Range
0 ... 999999 s
Default value
120 s
Measuring interval
Time interval for entering a measuring point in the measuring point list. Max.
number of entries?
712
Range
0.4 ... 999999 s
Default value
2.0 s
856
Range
0.1 ... 999999.0 s
Default value
2.0 s
Temperature
Measurement temperature
Temperature of the solution to be measured. If a temperature sensor is con-
nected, the temperature entered here is ignored.
Range
-20.0 ... 150.0 °C
Default value
20.0
5.6.5.8.4
CAL Cond - Conductivity standard
Tab: Method
▶ CAL Cond ▶ Properties... ▶ Conductivity standard
Command name
Name of the command.
Input
25 characters
Definition of the conductivity standard
Selection of the type as to how the data are to be defined for the conductivity
standard.