3 fixed bugs and problems, Fixed bugs and problems – Metrohm tiamo 2.2 Manual User Manual

Page 25

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

1 Introduction

tiamo 2.2

■■■■■■■■

11

Program part Configuration

■

Global variables, text templates and sample assignment tables can be
exported and imported.

■

When changing the titer, the values Date titer det., Titration
method
, User are additionally saved in the Audit Trail.

Program part Manual Control

■

If all stirrers are selected, no Stirring time for continuous operation can
be entered.

1.10.3

Fixed bugs and problems

General

■

When tiamo is closed before the login dialog appears, then an exception
may occur.

■

If having canceled after the entry of the licence update (tiamo 1.x to tiamo
2.x) and tiamo is normally started afterwards, an error message occurs.

Program part Workplace

■

Fast reading-in of barcodes can lead to the fact, that no more barcodes
can be read-in.

■

If a titration command is started via a SEND command in status Cond
not ok
, then no entry appears in the message list.

Program part Database

■

The data of a conductivity determination is not displayed with the menu
item Determinations

▶ Show calibration curve....

■

The blank value with a concentration calibration is not being recalculated
in the Recalculating.

■

Indexed method variables get lost during the Recalculating.

■

With fixed reports, in the selection list, the report Used titrants/solu-
tions
is missing.

■

The run test returns an exception when a WEIGH command is present
in the method.

Program part Method

■

DOS: For Titrinos, the stirrer mistakenly occurs on the tab General/
Hardware
.

■

In the method parameters report the field Tolerance is missing for CAL
Spec
.

Program part Configuration

■

When newly adding a Titrando, the firmware version is not being
checked.

■

The PREP warning does not appear for the 700 Dosino and 685 Dosimat.