5 Method
tiamo 2.2
■■■■■■■■
971
Range
0 ... 1.0
Default value
0.3
Slope
Minimum difference between the slope before and after the break point.
The smaller the difference, the more break-points will be found.
Range
0.0 ... 10.0
Default value
0.9
Smoothing factor
The higher the smoothing factor, the fewer endpoints will be found.
Range
2 ... 20
Default value
5
Window
Selection
Measured value | time | off
Default value
off
off
A range (window) can be defined on the measured value axis or on the
time axis. The break point evaluation will only be carried out in the
defined window. Only the first break point in the defined window will be
recognized.
Lower limit
Value for the lower limit of the window.
Time for the lower limit of the window.
Window = Measured value
Range
0.0000 ... 2000.0 mS/cm
Default value
0.0000 mS/cm
Window = Time
Range
0 ... 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.
Value for the upper limit of the window.
Window = Measured value
Range
0.0000 ... 2000.0 mS/cm
Default value
2000.0 mS/cm