5 Method
tiamo 2.2
■■■■■■■■
463
Lower limit pH
Lower limit for the measured value window.
Range
-20.000 ... 20.000
Default value
-20.000
Upper limit pH
Upper limit for the measured value window.
Range
-20.000 ... 20.000
Default value
20.000
EP criterion
Criterion for the recognition of endpoints. Endpoints whose discovered ERC
value is smaller than the value entered here will not be recognized.
Titrando, 855
Range
0 ... 200
Default value
5
EP recognition
Filter for the recognition of equivalence points:
Titrando, 855
Selection
first | greatest | last | ascending | descending
Default value
first
first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be recog-
nized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.