8 kft upol - additional evaluations, Additional evaluations – Metrohm tiamo 2.2 Manual User Manual

Page 787

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2

■■■■■■■■

773

Stabilizing time

Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.

Range

0 ... 999 s

Default value

0 s

Sample addition time

Waiting time between stop of conditioning and start of titration during
which the sample can be added.

Range

0 ... 99 s

Default value

0 s

Only start titration by a start command from a SEND command

on | off (Default value: off)

If this option is activated, then the titration will not be started until the Event
message
Start titration is sent to the command with a SEND command.

Note

Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titra-
tion will be ignored.

Show measured value during conditioning

on | off (Default value: off)

If this option is enabled, then the measured value will be shown in the sub-
window Live display of the program part Workplace, in addition to the
drift and the time.

5.6.3.5.3.8

KFT Upol - Additional evaluations

Tab: Method

▶ KFT Upol ▶ Properties... ▶ Additional evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of titration curves can be
activated and defined on this tab: