Metrohm tiamo 2.2 Manual User Manual
Page 650

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
636
■■■■■■■■
tiamo 2.2
as the measured value alterations between two measuring points are then
of the same order of magnitude as the noise.
Titrando, 855
Range
0.0001 ... 9.9999 mL
Default value
0.05 (slow), 0.10 (optimal), 0.20 (fast) mL
Titrino
Range
0.001 ... 9.999 mL
Default value
0.05 (slow), 0.10 (optimal), 0.20 (fast) mL
Dosing rate
Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the exchange unit or dosing unit
used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
Temperature
Temperature
Manually entered titration temperature. If a temperature sensor is connected
and the Temperature measurement is set on the General/Hardware
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
25.0 °C